A force sensor (also called a load cell or force transducer) measures tension or compression forces and converts them into an electrical signal.
DISA AG Sarnen supplies force sensors from Mark-10 and n-gineric for measurements from 0.5 N to 200 kN. This guide shows the criteria for choosing the right sensor. You will find all sensors in the Force sensor category.
The force sensors at a glance
Four criteria for your choice
1. Measuring range
- FS05: 0.5 N to 2.5 kN
- R03 and R04: 1 N to 500 N
- R01: 250 N to 50 kN
- R02: 500 N to 50 kN
- R07: 250 N to 7.5 kN
- ng-FTS-bi: 100 N to 50 kN
- ng-FTS-di: 1 kN to 200 kN
- FS06: 250 N to 2.5 kN
- R05: up to 2.5 kN
2. Accuracy
For Mark-10, depending on the series:
- ±0.15% of full scale: R01, R03, R05, R07, FS06
- ±0.2%: R04
- ±0.5%: R02
- FS05: ±0.1%
- n-gineric: ng-FTS-bi within 0.5%, ng-FTS-di within 2% of the measuring range
3. Design
- S-beam: R01
- Button sensor for compression: R02
- Miniature sensor: R04
- Fully enclosed: R05
- Ring compression sensor: ng-FTS-di
4. Connection
- R01 to R05 require a separately available Mark-10 indicator
- FS05, FS06 and R07 are designed for the F series universal testing machines
- n-gineric: own digital interface with USB, Ethernet or ng-Link
Plug & Test with Mark-10
With Mark-10 Plug & Test sensors, all configuration and calibration data are stored in the sensor. Sensors and indicators of the Plug & Test family are interchangeable.
Calibration and service in Sarnen
Our calibration laboratory in Sarnen is accredited to ISO/IEC 17025 (SCS 0083). We calibrate force with an SCS certificate up to 50 kN and with a factory report up to 200 kN.
Frequently asked questions
What is the difference between Mark-10 and n-gineric sensors?
The Mark-10 sensors R01 to R05 require a separately available Mark-10 indicator, FS05, FS06 and R07 are designed for the F series universal testing machines. The n-gineric sensors combine a strain gauge measuring element with an intelligent interface and communicate via USB, Ethernet or ng-Link.
Which sensors are available for very small forces?
FS05 from 0.5 N, and R03 and R04 from 1 N.
Can DISA calibrate the force sensors?
Yes, in our accredited laboratory in Sarnen (SCS 0083): with an SCS certificate up to 50 kN, with a factory report up to 200 kN.
Advice and quotation
Tell us the direction of force, measuring range, installation space and the connection you need. Our sales team will be happy to advise you: verkauf@disa.ch or +41 41 666 70 50.